Analisis Pengaruh Intensitas Pencahayaan Di Laboratorium Pengujian Terhadap Akurasi Pengujian
DOI:
https://doi.org/10.52759/inventory.v6i2.75Keywords:
Lighting Intensity, Testing Accuracy, Testing Laboratory, Correlation AnalysisAbstract
The testing laboratory is a high-precision testing room that reguires adeguate environmental conditions for testing needs.As a testing laboratory, it is important for the banten Indonesia Laboratory to control the laboratory's environmental conditions, one of which is lighting intensity.Inadeguate lighting intensity can reduce testing accuracy, Currently, the average lighting intensity in the banten Indonesia Laboratory is 370 lux, this value is below the regulatory standard Threshold value, which according to PERMENKES NO. 70 of 2016 for standard laboratories lighting intensity is 500 lux. This study aims to increase lighting according to SNI 7062:2019 and testing accuracy in the Banten Indonesia Laboratory and to determine how strong the relationship betwen lighting intensity and testing accuracy is using correlation analysis with SPSS V.27. The results of this study indicate that lighting intensity has increased to 489 lux and the average accuracy has increased by 7.5 % after changing the lamp,in addition,the relationship between lighting intensity and testing accuracy strong ,this is evidenced by the result of the correlation analysis obtained a significance value of 0.620 and 0.627 .
Downloads
References
Nasional, B. S. (2017). Standar Nasional Indonesia Persyaratan umum kompetensi laboratorium pengujian dan kalibrasi General requirements for the competence of testing and calibration laboratories. In IEC (Vol. 17025). www.bsn.go.id
Putri, S. I., & Sudarti, S. (2022). Analisis Intensitas Cahaya di Dalam Ruangan dengan Menggunakan Aplikasi Smart Luxmeter Berbasis Android. Jurnal Materi Dan Pembelajaran Fisika, 12(2), 51. https://doi.org/10.20961/jmpf.v12i2.51474
Rachmat, A. N., Santoso, M. Y., & Zakaria, R. (2023). Analisis Nilai Pengukuran Penerangan pada Laboratorium Ergonomi sesuai Permenaker No 05 Tahun 2018. G-Tech: Jurnal Teknologi Terapan, 7(2),422–428. https://doi.org/10.33379/gtech.v7i2.1998
Syamsi, W. A., Setianto, N., Saepuloh, A., & Suyono, A. M. (2024). Analisis Intensitas Cahaya Di Laboratorium Pt X Metode Sni 7062:2019. Jurnal Ilmiah Teknik Dan Manajemen Industri Jurnal Taguchi,4(1),163–170. https://doi.org/10.46306/tgc.v4i1
Tegar, F., Alawiyah, R., Hikmah Dilla Jannah, K., Adi Jaya Pohan, A., Mardyanti Purba, N., Hasanah, N., Anggraini Nasution, D., & Niswati Utami, T. (2023). Pengaruh Intensitas Paparan Cahaya Terhadap Kelelahan Mata Pada Pekerja Home Industri. Ilmiah Mahasiswa Pendidikan Sejarah, 8(3),2035–2045. https://doi.org/10.24815/jimps.v8i3.25616
Yudha Asmara, I., Garnida, D., & Hilmia, N. (2023). Korelasi Intensitas Cahaya Dengan Karakteristik Suara Ayam PeluNG (Correlation Of Light Intensity and Sound Characteristics of Pelung Chickens). 49, 2024–2376.